Explanation
BACKGROUND: The Division of Police needs to purchase a Fourier Transform Infrared Spectrometer (FTIR) with an Attenuated Total Reflectance accessory system to be used for the identification of controlled substances at the Police Crime Laboratory. The prompt acquisition of this instrument will assist greatly in drug identification cases. The Fourier Transform Infrared Spectrometer instrument is used to identify the presence of controlled substances seized as evidence examples.
Bid Information: The Purchasing Office conducted a formal bid #SA003728, which opened on October 28, 2010. Four companies responded: PerkinElmer Health Sciences, Inc. (WHT), Varian, Inc. (PHC), Thermo Electron North American LLC (not contract compliant), and Mantua Chemical Consulting, Inc. (not contract compliant) with bid prices as follows:
PerkinElmer -$57,453.20 Thermo Electron NA LLC -$33,024.77
Mantua Chemical -$29,370.00
Varian -$28,172.00
The Division of Police recommends the bid submitted by PerkinElmer Health Sciences, Inc., as the most responsive, responsible and best bid meeting all specifications. PerkinElmer Health Sciences Inc., is the only vendor who met the bid specifications in the entirety. The three low bidders failed to conform to the specifications based upon the following information.
The following specifications were not met by Mantua Chemical Consulting, Thermo Electron North America, and Varian: 3.49: The spectrometer must utilize a rotary interferometer and require no scanning mirror dynamic alignment. Dynamically aligned and corner cube design interferometers are not acceptable.
The three low bidders all bid the dynamically aligned interferometers.
Specification 3.6.8 - The supplied computer shall have an Intel Pentium 4 processor (2.8 GHz or faster), with a minimum 1 GB RAM, DVD/RW/CD-RW Drive, 19" Flat Panel Color Monitor, and dual hard disk storage (minimum 80GB total). All three low bidders specifi...